YL4520 spectrophotometer is a color measuring instrument with 45/0 optical geometry(45 ring-shaped illumination, vertical viewing), to achieve precise non-contact measurement between the measuring probe and the sample like liquids, sauces, powders, etc, which is developed by 3nh independently with independent intellectual property.
YL4520 non-contact benchtop spectrophotometer can be applied for accurate color analysis and transmission in laboratories and R & D industries; it can also be used for accurate color measurement and quality control of automated production lines; it is widely used in cosmetics, fruits and vegetables, food hygiene, plastic electronics, paint and ink, printing, ceramics and other industries. Its unique innovative design can not only provide non-contact measurement scheme directly from the production line, but also ensure stable and high-precision measurement results.
YL4520 non-contact spectrophotometer will not touch the sample in the measurement process, to prevent damage to the sample, but also to avoid cross contamination.
PRODUCT FEATURES
1. non-contact, 45 / 0 geometric optical structure to measure the reflectance and chromaticity of objects
2. The movable measuring head moves up and down according to the actual height of the measured object
3. Large touch screen measurement interface, real-time display of measurement data, to realize more measurement functions
4. Multiple measurement modes (sample, quality control, continuous statistical mode) can be selected to meet the personalized needs
APPLICATION INDUSTRY
45/0 non- contact benchtop spectrophotometer can achieve precise non- contact measurement between the measuring probe and the sample like liquids, sauces, powders, etc. It can be applied for accurate color analysis and transmission In laboratories and R & D industries; It can also be used for accurate color measurement and quality control of automated production lines; It is widely used in cosmetics, fruits and vegetables, food hygiene, plastic electronics, paint and ink, printing, ceramics and other industries.
Specification
Model | YL4520 |
Optical Geometry | 45/0 method(45 ring-shaped illumination, vertical viewing);Conforms to CIE No.15,GB/T 3978,GB 2893,GB/T 18833,ISO7724-1,ASTM E1164,DIN5033 Teil7,GB 2893,GB/T 18833 |
Features | 45/0 non- contact benchtop spectrophotometer can achieve precise non- contact measurement between the measuring probe and the sample like liquids, sauces, powders, etc. It can be applied for accurate color analysis and transmission In laboratories and R & D industries; It can also be used for accurate color measurement and quality control of automated production lines; It is widely used in cosmetics, fruits and vegetables, food hygiene, plastic electronics, paint and ink, printing, ceramics and other industries. |
Light Source | Full spectrum LED light |
Spectrophotometric Mode | Concave Grating |
Sensor | 256 Image Element Double Array CMOS Image Sensor |
Wavelength Range | 400~700nm/10nm output |
Reflectance Range | 0~200% |
Measuring Aperture | Φ20mm |
Non-contact Distance | 7.5mm |
Sample Height | 0~150mm |
Distance Adjustment Method | Manual adjustment(test height can be stored) |
Measuring mode | Sample mode/QC Mode/Continuous statistics mode |
Locating Method | Camera Locating |
Color Space | CIE LAB,XYZ,Yxy,LCh,CIE LUV,HunterLAB |
Color Difference Formula | ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00,ΔE(Hunter) |
Other Colorimetric Index |
WI(ASTM E313,CIE/ISO,AATCC,Hunter),YI(ASTM D1925,ASTM 313),TI(ASTM E313,CIE/ISO), Metamerism Index MI, Staining Fastness, Color Fastness, Color Strength, Opacity |
Observer Angle | 2°/10° |
Illuminant | D65,A,C,D50,D55,D75, F1,F2(CWF),F3,F4,F5,F6, F7(DLF),F8,F9,F10(TPL5), F11(TL84),F12(TL83/U30) |
Displayed Data | Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, Chromaticity Graph, Color Simulation, PASS/FAIL Result, Color Offset |
Measuring Time | About 1.5s |
Data Storage | Sample mode + QC mode: 30,000; Continuous statistics mode: 10,000 |
Repeatability |
Spectral reflectance: Standard deviation within 0.08% Chromaticity value: Within ΔE*ab 0.05 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration) |
Inter-instrument Error | Within ΔE*ab 0.25(Average for 12 BCRA Series II color tiles) |
Measurement Mode | Single Measurement, Average Measurement(2 ~ 99 times) |
Dimension | 330*250*370mm |
Weight | About 10 kg |
Power Supply | DC 24V, 3A Power Adapter |
Illuminant Life Span | 5 years, more than 3 million times measurements |
Display Screen | 7" TFT Capacitive Screen-touch Display |
Data Port | USB/RS-232 |
Language | Simplified Chinese,Traditional Chinese,English |
Working Environment | Temperature: 0~40℃; Humidity: 0~85% (No Condensation); Altitude < 2000m |
Storage Environment | Temperature: -20~50℃; Humidity: 0~85% (No Condensation) |
Standard Accessory | Power Adapter, User Manual, USB cable, Standard Calibration Board, Black Calibration Cavity, |
Notes | The specifications are subject to change without notice. |
With 12 rigorous quality inspection processes in place, we ensure a high yield rate of up to 98% for batch processed precision parts.
FAQs About Spectrophotometer
Yes, we can offer quality certification issued by our lab free of charge, or issued by third-party international inspection Organizations with extra charge, please contact us for confirmation.
After receiving NH310/NH300, you have to perform white and black calibration manually when first starting.
a. When the colorimeter is disconnected to the PC, "Connection Timeout" will occur. At the moment, you have to check whether the USB cable is well connected the colorimeter and the PC. Then you can re-plug the USB to try out.
b. After ensuring the colorimeter is connected to the PC with USB cable, turn on the colorimeter and enter the main menu. Select "Comm" and press "Enter" key to start communication.
c. Turn off CQCS3 software, and reopen it.
Threenh colorimters are used across various industries to measure, analyze, and control color accuracy and consistency:
Food and Beverages: Monitor color quality in food products to meet industry standards.
Plastics and Polymers: Ensure color uniformity in plastic components (e.g., casings, buttons, panels) during manufacturing.
Chemicals and Pharmaceuticals,Cosmetics: Industry-specific accessories to attain the highest safety protocols, Ensure uniform color in pharmaceutical products for quality and brand recognition;Develop and match colors for cosmetics (e.g., lipsticks, creams, powders) and ensure batch consistency.
Textiles and Apparel: Monitor color consistency in fabrics during dyeing, printing, or finishing processes; ensure color consistency in fabrics and garments.
Building materials, Paints, and Coatings: Detect color deviations in coatings on metal, wood, or plastic surfaces (e.g., automotive paints, furniture finishes). Maintain color accuracy in building materials, paints, and coatings.
Printing and Packaging: Validate color accuracy in printed materials (e.g., brochures, labels, packaging) against design proofs, maintain consistency across different printing technologies.
By addressing color-related quality control challenges, 3NH colorimeters help industries maintain brand consistency, meet regulatory standards, and enhance customer satisfaction.
a. It’s needed to perform manual calibration for NH310 when first starting.
b. In the subsequent starting, NH310 will perform auto calibration at startup. No need to perform manual calibration.
c. Suggestion: it’s best to perform manual calibration once a week or when NH310 measured data is not accurate.